User Tools

Site Tools


data:data_analysis_manual

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revision Previous revision
Next revision Both sides next revision
data:data_analysis_manual [2024/01/17 16:26]
Peter Young [Units and uncertainties]
data:data_analysis_manual [2024/01/23 19:27]
eric buchlin [Burn-in Effects] corrected files published in data release 4.0
Line 95: Line 95:
 ==== Burn-in Effects ==== ==== Burn-in Effects ====
  
-The SPICE detectors have microchannel plates (MCPs). Over time, the sensitivity of MCPs decreases at the locations of strong emission lines. This effect is referred to as "​burn-in"​ and results in a flat-top or "​self-reversal"​ to the line profile. This has already been noticed for the C III 977 line. As of November 2023internal ​level-2 files are corrected for burn-in for six lines (files planned to be part of release 4.0). The effects of burn-in continue to be monitored, particularly for weaker lines for which a correction is not currently available.+The SPICE detectors have microchannel plates (MCPs). Over time, the sensitivity of MCPs decreases at the locations of strong emission lines. This effect is referred to as "​burn-in"​ and results in a flat-top or "​self-reversal"​ to the line profile. This has already been noticed for the C III 977 line. From data release 4.0, level-2 files are corrected for burn-in for six lines. The effects of burn-in continue to be monitored, particularly for weaker lines for which a correction is not currently available.
  
 The lines that are being corrected (as of 2023-11) are: N IV 76.5 nm, Ne VIII 77.0 nm, Ne VIII 78.0 nm, C III 97.7 nm, Lyman-beta 102.6 nm, and O VI 103.2 nm. The lines that are being corrected (as of 2023-11) are: N IV 76.5 nm, Ne VIII 77.0 nm, Ne VIII 78.0 nm, C III 97.7 nm, Lyman-beta 102.6 nm, and O VI 103.2 nm.
data/data_analysis_manual.txt · Last modified: 2024/03/29 14:21 by eric buchlin